The five points
AccessRA uses five working points across the gauge span: 0%, 25%, 50%, 75% and 100%. For a 0–1000 psi gauge, that is 0, 250, 500, 750 and 1000 psi.
For another range, calculate the points from that span rather than copying the training pressures.
AccessRA Academy · Workshop tool
Hydraulic deadweight tester
Existing AccessRA Academy workshop-tool record reused here; no duplicate media upload.
Example: Hydraulic pressure balance / deadweight tester
Try it: complete the upscale run
Apply each reference point in order, allow the bench and pointer to stabilise, then read and record the gauge. Do not adjust during the As-Found sequence.
Calculate the error
Error = indication − reference. To express it against span: %FS error = error ÷ full-scale span × 100.
Compare the result with the tolerance specified for that gauge/job. The 0–1000 psi Class 1.0 AccessRA training case uses ±10 psi.
Do not stop at the upscale run
A complete calibration job may require a controlled 100→75→50→25→0% downscale run, diagnosis, correction only when justified, and a fresh full As-Left verification. Preserve the original As-Found data.